The Agilent 710 Series ICP-OES is a cutting-edge, high-performance instrument designed for elemental analysis across a wide range of industries. Featuring advanced technology and enhanced sensitivity, the 710 Series delivers accurate and reliable results with unmatched precision. Its user-friendly interface and intuitive software allow for easy operation, while the robust design ensures reliable performance in demanding laboratory environments. The Agilent 710 Series is engineered for flexibility, with customisable configurations and compatibility with a wide variety of sample formats, making it a versatile tool for both routine analysis and challenging research applications.
Key Features of the Agilent 710 Series ICP-OES:
- Temp control – It is recommended that the ambient temperature of the laboratory be between 20 and 25 °C (68 and 77 °F) and be held constant to within ±2 °C (±3.6 °F) throughout the entire working day
- Wide wavelength range: 164 to 847 nm, covering a broad range of elements.
- Plasma:
- Plasma temperature: 10,000 K (typical).
- The work coil operates at 1,500 V RMS and about 40 MHz
- Plasma torch: High efficiency and robustness for analysing various sample types.
- Includes:
- ASX-520 Autosampler
- Van der Heijden chiller
- Full computer system with software and monitor
ASX-520 Autosampler Specifications:
- Power Requirements: 100 – 240 VAC +/- 10%, 50/60 Hz
- Tray: 4 rack capacity
- Racks: 90 x 7 ml, 60 x 14 ml, 40 x 20 ml, 24 x 30 ml, 21 x 50 ml
- Height: 25 cm (9.8”)
61 cm (24”) – with sample probe - Width: 52 cm (20.5”)
- Depth: 48.2 cm (19”)
- Weight: 10.5 kg (23 lbs)
Van der Heijden Chiller Specifications:
- Temperature range: 16°C to 22°C
- Cooling power (at 20°C): 600 watt
- Pump capacity max. 0.4 bar pump: 28 l/min
- Pump capacity max. 3.5 bar pump: 10 l/min
- Dimensions (WxDxH): 360 x 470 x 590 mm
- Weight: 52 kg